Magstim FAQ 2
2. How many times can the coils stimulate before they go over-temperature?
Please refer to the following table to find the maximum number of stimulations for each specific coil.
|
Coil |
Maximum Number of Stimulations at 50% Power |
Maximum Number of Stimulations at 80% Power |
Maximum Number of Stimulations at 100% Power |
|
High Power 90mm |
>505 |
258 |
143 |
|
70mm Medium |
256 |
98 |
63 |
|
50mm Small |
281 |
104 |
65 |
|
Double 70mm |
338 |
108 |
56 |
|
Double Cone 110mm |
>584 |
>584 |
584 |
|
Double 25mm |
146 |
55 |
40 |
|
Double 70mm Flat |
633 |
170 |
121 |
|
Double 70mm Branding Iron |
1023 |
176 |
123 |
|
Double 70mm Placebo |
>984 |
>984 |
984 |
|
2nd Generation 90mm |
505 |
260 |
145 |
|
2nd Generation Double 70mm |
340 |
110 |
60 |
The "maximum number of stimulations" columns relate to the number of stimulations that can be achieved with a Magstim 200 or Magstim 200², at an ambient temperature of 20oC to 25oC, before the coil exceeds its permitted operating temperature.