Monopolar Probe
- Single Use
- Monopolar Probe
- Sterile packed in boxes of 10
- P/N 3602-00

This stimulating probe is intended for stimulating across the bulk of a tumour to determine whether a nerve is lying behind a structure, and is not designed for accurate and selective nerve stimulation. However, accuracy can be improved by inserting the ground electrode as close as possible to the surgical area, and by using the minimum current that will stimulate the nerve.
A bend may be placed in the shaft of the probe for use under a microscope.
Please note: a ground electrode, P/N 1705-00, is also required when using this probe.
For further information, queries or feedback, please contact neurosign.info@magstim.com