Magstim DC Stimulator

Transcranial direct current stimulation (tDCS) involves applying weak electrical currents to the head, to generate an electromagnetic field which modulates the activity of brain neurons. tDCS has been shown to selectively modulate neuronal excitability and can be used in conjunction with Magstim transcranial magnetic stimulation (TMS), fMRI or centrally acting drugs. It is being investigated as a treatment for a variety of conditions such as stroke recovery, depression and migraine. tDCS should not be confused with electroconvulsive therapy (ECT) or TMS. tDCS is a technique that does not cause muscle twitches, seizures or other unwanted effects associated with older forms of electrical stimulation.

The Magstim eldith DC Stimulator (now available) is an ideal device for research in this field, being both portable and durable, with clear displays and easy to use controls.  The variety of settings give researchers the scope that they need to conduct their studies.

Features

  • Microprocessor controlled constant current source.Magstim tDCS Stimulator
  • High safety standard through multistage monitoring of output current.
  • 1 channel (anodal and cathodal stimulation possible).
  • Adjustable current up to 5,000µA (±1 %).
  • Adjustable application time up to 30 min.
  • Modes of stimulation:
    • Single (continuous stimulation, adjustable fading in and fading out).
    • Pulses (cyclic turning on/off of the stimulation, adjustable pulse width and interval).
  • "Study Mode" (optional), for fully blinded operation of real and "pseudo" stimulation, encoded via code list of 200 codes, independently adjustable settings (can be saved to avoid accidental changes of study parameters).
  • External trigger input (optional), e.g. for remote controlling.
  • Please Note: currents of more than 2,000µA and application times of more than 20 min are for research purposes only.

For regulatory issues information in the USA please click here.

For futher information, queries or feedback, please contact magstim.info@magstim.com

CAUTION: Investigational Device. Federal or (United States) law limits device to investigational use.

Specification

  • 135mm x 225mm x 55mm (W x D x H); weight 0.8kg (incl. batteries).
  • 12bit D/A conversion.
  • Time resolution 5ms (sample rate 200sps).
  • Current variance during stimulation < 0.02 %
  • Power consumption approx. 0.2-1.5W (depends on backlight brightness and applied current).
  • Power supply via built-in rechargeable batteries, incl. external charger.
  • alphanumeric display with backlight, membrane keyboard with 4 keys.
  • Touch-proofed safety connectors acc. to DIN 42802-2 (ø 1.5mm)
  • Approx. 6h stimulation time @ 1mA, approx. 7h for fully recharging.
  • Trigger Module for safe connecting of external trigger (optional).